Reliability of a computer-aided manual procedure for segmenting optical coherence tomography scans
Donald C Hood1, Jungsuk Cho, Ali S Raza
1Department of Psychology, Columbia University, New York, New York 10027, USA. dch3@columbia.edu
Summary
Manual segmentation of frequency-domain optical coherence tomography scans for glaucoma studies shows good reliability. Trained operators achieve consistent results, ensuring accurate retinal layer thickness measurements.
Area of Science:
- Ophthalmology
- Medical Imaging
- Biomedical Engineering
Background:
- Accurate measurement of retinal layer thickness is crucial for diagnosing and monitoring glaucoma.
- Frequency-domain optical coherence tomography (FD-OCT) is a key imaging modality for glaucoma assessment.
- Manual segmentation of FD-OCT scans requires consistent and reliable performance from operators.
Purpose of the Study:
- To evaluate the consistency of manual segmentation of retinal layers in FD-OCT scans.
- To assess both within-operator (test-retest) and between-operator agreement.
- To determine the reliability of computer-aided manual segmentation for glaucoma research.
Main Methods:
- Four trained individuals manually segmented FD-OCT scans from glaucoma patients and healthy controls using a computer-aided tool.
- Scans were segmented twice by each operator with at least a one-month interval to assess within-operator reliability.
- Concordance correlation coefficients and mean absolute differences in border locations were used to quantify agreement.
Main Results:
- Good within-operator reliability was observed, with mean concordance correlations ranging from 0.978 to 0.999.
- Mean differences in border locations (MEAN{mean( ΔLBL )}) ranged from 1.6 to 4.7 micrometers.
- Good between-operator agreement was also found, with concordance correlations between 0.992 and 0.999, and MEAN{mean( ΔLBL )} values from 1.9 to 4.0 micrometers.
Conclusions:
- Trained human segmenters demonstrate high reliability in manual border segmentation of FD-OCT scans.
- Manual segmentation provides a dependable method for measuring retinal layer thickness in glaucoma studies.
- The consistency of this method supports its use in clinical research for glaucoma assessment.