Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array
Dayananda Shamurailatpam Sharma1, Vaibav Mhatre, Malhotra Heigrujam
1Department of Radiation Oncology, Kokilaben Dhirubhai Ambani Hospital and Medical Research Institute, Mumbai-400053, India. dayananda.sharma@relianceada.com
Abstract:
Portal dosimetry (PD) was performed for 181 fields from 14 IMRT plans of various clinical sites at gantry zero and source-to-detector distance (SDD) of 100 cm. PD was realized using aSi1000 electronic portal imaging device (EPID) and portal dose prediction (PDP) algorithm implemented in Eclipse treatment planning system (TPS). Agreement of PDP predicted and EPID measured photon fluence/dose distribution were evaluated using gamma (γ) index set at 3% at 3 mm distance to point agreement (DTA). Three gamma scaling parameters, maximum γ (γ(max)), average γ (γ(avg)) and percentage of points with γ ≤ 1 (γ% ≤ 1) were estimated for each field. An independent measurement was carried out using MatriXX 2D ion chamber array with detector plane at 100 cm and γ(max), γ(avg) and γ% ≤ 1 were estimated using OmniPro IMRT analyzing software. Effect of extended SDD and gantry rotation on portal dosimetry outcome was also investigated for another 45 IMRT fields. PDP predicted and EPID measured photon fluence agrees well with overall mean values of γ(max), γ(avg) and γ% ≤ 1 at 2.02, 0.24 and 99.43%, respectively. γ(max) value was lower in 15 MV compared to 6 MV IMRT plan. Independent verification using MatriXX showed comparable overall mean values of γ(avg) and γ% ≤ 1 at 0.25 and 99.80%. However, in all plans, MatriXX showed significantly lower γ(max) (p < 0.05) with an overall mean value of 1.35. In portal dosimetry, compared to gamma values at 100 cm SDD, γ(max), γ(avg) and γ% ≤ 1 values improve from a mean of 0.16, 0.03 and 0.26 at 110 cm SDD to 0.35, 0.05 and 0.29 at 140 cm SDD. PD outcome was independent of gantry rotation. In conclusion, both MatriXX 2D ion chamber array and portal dosimetry showed comparable results and can be use as an alternative to each other for relative photon fluence verification.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014