Comparison of two new generation pulse oximeters during emergency ambulance transportation
Ulrike Weber1, Elvira Tomschik, Irene Resch
1Department of Anesthesiology, General Intensive Care and Pain Control, Medical University of Vienna, Austria. ulrike.weber@meduniwien.ac.at
Objectives:
We wanted to test whether there is a difference between the total number and duration of malfunctions and a correlation between the oxygen saturation and pulse rate values of two new generation pulse oximeters (Masimo 'Radical 7' and Nellcor 'N 600') during emergency ambulance transportation.
Methods:
Patients were monitored with two pulse oximeters ('Radical 7' and 'N 600') on different randomly selected fingers of the same hand during transportation. Data of both devices were recorded continuously by a laptop computer.
Results:
Fifty-two patients with signs of peripheral vasoconstriction (including 22 patients with a blood pressure ≤100/60) were included. There were 0.21 ± 0.72 (0-4) malfunctions per patient lasting for a mean 113.55 ± 272.55 s in the 'Radical 7' and 0.13 ± 0.49 (0-3) malfunctions per patient with a mean duration of 301.0 ± 426.58 s in the 'N 600'. Oxygen saturation and pulse rate values correlated significantly [r² = 0.9608 (SpO₂), r² = 0.9608 (pulse rate)] between the devices and showed a bias of -0.177770 (SpO₂) and 0.310883 (pulse rate) with a standard deviation of 1.68367 (SpO₂) and 4.46532 (pulse rate) in a Bland-Altman test.
Conclusion:
Although number and duration of malfunctions did not differ significantly between the devices, they showed a very low number of malfunctions even in hypotensive patients with peripheral vasoconstriction. Oxygen saturation correlated significantly in the two devices investigated at 49.409 time points. In addition, pulse rate also correlated significantly.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...