Low versus high gas flow rate for respiratory support of infants at birth: a manikin study
Arjan B te Pas1, Kim Schilleman, Mirjam Klein
1Division of Neonatology, Department of Pediatrics, Leiden University Medical Center, Leiden, The Netherlands. a.b.te_pas@lumc.nl
Background:
Neonatal resuscitation guidelines do not specify the gas flow rate during mask ventilation.
Aim:
Investigating the effect of gas flow rates on pressures, volumes delivered and mask leak.
Methods:
Flow 5 and 10 liters/min were tested. In study part 1, pressure ranges were measured when ventilating an intubated manikin with a Neopuff®. In study part 2, pediatric staff mask-ventilated a manikin (peak inflation pressure (PIP) 30 cm H(2)O, positive end expiratory pressure (PEEP) 5 cm H(2)O). We measured pressures, expired tidal volume (V(Te)) and mask leak.
Results:
Study part 1:an intubated manikin was ventilated with flow 5 versus 10 liters/min: range in PEEP was 0.4-3.6 and 2-14 cm H(2)O, respectively, maximum PIP was 73 cm H(2)O with both flow rates. Study part 2: when mask ventilation was given with flow 5 versus 10 liters/min: leak decreased (24% (8-85) vs. 80% (34- 94); p < 0.0001), V(Te) increased (6.7 (5.1-7.8) vs. 4.7 (2.4-7.0) ml; p < 0.001), PEEP decreased (3.1 (0.8) vs. 3.7 (0.7) cm H(2)O; p < 0.001), PIP was similar (28.1 (2.7) vs. 28.0 (2.3) cm H(2)O; NS). Large leaks decreased V(Te) and PEEP during both flow rates, PIP only with flow 5 liters/min.
Conclusion:
A low flow rate during neonatal mask ventilation may be a good alternative approach in reducing mask leak, provided that inflation time and flow rate warrants set pressures. Only large leaks seem to influence delivered pressures and volumes. Before resuscitation guidelines are advised, more studies on gas flow rates are needed.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:25Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Respiratory Volumes and Capacities
Respiratory Volumes and Capacities I
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...