Background:
The aim of this study was to investigate the imaging capabilities of recent hand-held ultrasound scanners.
Methods:
Three hundred forty-nine patients were scanned with hand-held ultrasound (HAND) and high-end echocardiography (HIGH). Segmental endocardial border delineation was scored (2 = good, 1 = poor, 0 = invisible) to describe image quality. Assessments of left ventricular (LV) dimensions, regional and global LV function, and grades of valve disease were compared.
Results:
The mean endocardial visibility grades were 1.6 ± 0.5 with HAND and 1.7 ± 0.4 with HIGH (P < .01). Regional wall motion was scored very similarly (κ = 0.73, P < .01). Ejection fraction assessment (bias = 1.8%, 1.96 × SD = 8.3%) and LV measurements (r = 0.99, P < .01; interventricular septum: bias = 0.91 mm, 1.96 × SD = 2.1 mm; LV end-diastolic diameter: bias = 0.5 mm, 1.96 × SD = 4.1 mm; LV posterior wall: bias = 0.61 mm, 1.96 × SD = 2.4 mm) showed negligible deviations. No pericardial effusion or valve stenosis was missed. Regurgitations missed by HAND were all graded "minimal" on HIGH. Regurgitations were mildly overestimated by HAND. Overall concordance for detection of regurgitations was very good (κ = 0.9, P < .01).
Conclusions:
Handheld echocardiography was feasible and missed no relevant findings. Given the future implementation of spectral Doppler capabilities, this handheld scanner can safely be used in clinical routine.
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Acute Coronary Syndrome III: Diagnostic Studies