Multicenter user evaluation of ACCU-CHEK® Combo, an integrated system for continuous subcutaneous insulin infusion
David Kerr1, Roel P L M Hoogma, Andreas Buhr
1Centre of Postgraduate Medical Research and Education, School of Health and Social Care, Bournemouth University, Dorset, United Kingdom. dkerrmd@gmail.com
Background:
The aim of this study was to evaluate a newly developed system for insulin delivery incorporating a multifunctional blood glucose meter and a remotely controlled insulin pump (ACCU-CHEK® Combo system) in established pump users with type 1 diabetes. The technology was assessed both from device performance and subject usability perspectives.
Method:
A multicenter, prospective, single group study was carried out in five centers in the Netherlands and four centers in the United Kingdom for more than 6 months. The study was divided into two phases: Phase 1 (4 weeks) for device validation purposes and phase 2 (22 weeks) to observe the impact of the system on metabolic control, patient satisfaction [using the Diabetes Treatment Satisfaction Questionnaire (DTSQ)] and device safety.
Results:
Eighty subjects completed the planned study period. There were no unexpected device errors. Treatment satisfaction was high at baseline and further increased to study end (DTSQ change version: sum score, 10.6±7.2; scale score range, -18 to +18, p<0.0001). Hemoglobin A1c improved continuously over time, from 7.9% (±0.9%) to 7.7% (±0.8%) at month 3 (p<0.001) and 7.6% (±0.8%) at month 6 (p<0.0001). The frequency of severe hypoglycemia was 0.08 per patient years. There was no case of ketoacidosis.
Conclusions:
The new system was evaluated by experienced continuous subcutaneous insulin infusion users as safe in daily practice and associated with favorable treatment satisfaction and a modest improvement in glycemic control.
More Related Videos
08:22Combined Intravital Microscopy and Contrast-enhanced Ultrasonography of the Mouse Hindlimb to Study Insulin-induced Vasodilation and Muscle Perfusion
Published on: March 20, 2017
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into rapid-acting...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Integrated Healthcare System
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices