Published on: January 17, 2011
How can a single rescuer adequately deliver tidal volume with a manual resuscitator? An improved device for
Yong Chul Cho1, Sung Wook Cho, Sung Pil Chung
1Department of Emergency Medicine, Chungnam National University Hospital, Daejeon, Republic of Korea.
Objectives:
A bag-valve mask (BVM) device is used as one of the first-line pieces of equipment in emergency situations. However, cardiopulmonary support providers do not recognise the exact tidal volume during procedures, and squeezing methods of BVM may not deliver the same tidal volume each time. To supply a regular and sustained tidal volume, adequate finger points were marked on the surface of a BVM.
Methods:
In this study, a total of 83 volunteers participated and practised conventional BVM and volume-marked bag-valve mask (VBVM) procedures. The VBVM is simply a conventional BVM with an imaginary axis grid, drawn to guide the placement of the fingers. The VBVM method provides a constant volume of approximately 500-600 ml; the bag is squeezed until the thumb and the middle finger touch slightly. The results were then statistically analysed.
Results:
The tidal volume delivered by the studied VBVM method is more accurate than the conventional BVM method (421.87±95.19 ml vs 534.21±24.22 ml, p<0.001). There was no statistical correlation except age between the results and the participants' training level or physical characteristics in the study.
Conclusions:
As the conventional BVM method cannot deliver a regular and sustained tidal volume, the authors invented the VBVM method. This method delivered a volume of 500-600 ml with more stability each time, which can improve the outcome of emergency patients.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques