Related Experiment Video
Updated: Jun 6, 2026

Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
[Application of a new catheter to be determining cardiac output through cor sinistrum with thermodilution]
Fei Huang1, Xiao-Ling Li, Li-Min Zheng
1Department of Anesthesiology, Shenzhen Hospital, Peking University, Shenzhen 518035, China.
Aim:
To introduce a new device and catheter to be determining cardiac output through cor sinistrum with thermodilution.
Methods:
Own control was introduced in eight dogs. A pulmonary thermodilution catheter was used through the cor dextrum of the dogs; The new-design catheter was used through artery, and there were two situs to fix:its ahead in left atrium and the ahead near to aortic root. Determine cardiac output (CO) under four circumstances: controlled respiration, apnoea, in controlled hypotension, after controlled hypotension.
Results:
There was no significant difference among the parameters (CO and cardiac index) of all circumstances. The CO determined by the pulmonary thermodilution catheter and those by the new-design catheter in two situs were positive correlation, and the gamma were 0.986, 0.989; likewise, the cardiac index (CI) was positive correlation, and the gamma were 0.983, 0.985. The CO and CI by the new-design catheter between two situs were positive, and the gamma were 0.992, 0.988.
Conclusion:
The parameters by the new-design catheter and those by the pulmonary thermodilution catheter were concordant. In comparison with the pulmonary thermodilution catheter, the new-design catheter had simple device and could be easily operated.
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...