What does it take to start an oropharyngeal fire? Oxygen requirements to start fires in the operating room
1Pediatric Otolaryngology, University of Texas-Houston School of Medicine, United States. sohamroy0@gmail.com
Introduction:
Airway fires are a well-described and potentially devastating complication of oropharyngeal surgery. However, the actual factors required to ignite the fire have never been well-delineated in the medical literature. In this study, we used a mechanical model to assess the oxygen parameters necessary to cause an oropharyngeal fire.
Methods:
An electrosurgical unit (Bovie) was grounded to a whole raw chicken and a 6.0 endotracheal tube (ETT) was inserted into the cranial end of the degutted central cavity. Oxygen (O(2)) was then titrated through the ETT tube at varying concentrations, with flow rates varying from 10 to 15L/min. Electrocautery (at a setting of 15W) was performed on tissue in the central cavity of the chicken near the ETT. All trials were repeated twice to ensure accuracy. Positive test results were quantified by the time required to obtain ignition of any part of the mechanical setup and time required to produce a sustained flame. A test was considered negative if no ignition could be obtained after four minutes of direct electrocautery.
Results:
At an O(2) concentration of 100% and a flow rate of 15L/min, ignition with a sustained flame was obtained between 15 and 30s after initiation of electrocautery. At 100% O(2) at 10L/min, ignition was obtained at 70s with immediate sustained flame. At an O(2) concentration of 60%, ignition occurred at 25s and sustained fire after 60s. At an O(2) concentration of 50% ignition with a sustained flame occurred between 128 and 184s. At an O(2) concentration of 45%, neither ignition nor sustained flames could be obtained in any trial.
Conclusions:
Operating room fires remain a genuine danger when performing oropharyngeal surgery where electrocautery is performed in an oxygen-enriched environment. In our study, higher O(2) flow rates with higher FiO(2) correlated with quicker ignition in the chicken cavity. A fire was easily obtained when using 100% O(2); as the O(2) concentration decreases, longer exposure to electrocautery is required for ignition. Below 50% O(2) we were unable to obtain ignition. Our study is the first to examine the relative risk of ignition and sustained fire in a mechanical model of oropharyngeal surgery. Decreasing the fraction of inspired O(2) (FiO(2)) to less than 50% may substantially decrease the risk of airway fire during oropharyngeal surgery.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...