Related Experiment Video
Updated: Jun 6, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Radiation dose and image quality for paediatric interventional cardiology systems. A national survey in Chile
Carlos Ubeda1, Eliseo Vano, Patricia Miranda
1Clinical Sciences Department, Health Sciences Faculty and CIHDE, Tarapaca University, Arica, Chile. cubeda@uta.cl
Abstract:
Radiation dose and image quality for paediatric protocols in all five X-ray fluoroscopy systems used for interventional cardiology procedures existing in Chile have been evaluated. Entrance surface air kerma (ESAK) and image quality using a test object (TO) and polymethyl methacrylate (PMMA) phantoms have been measured for the typical paediatric patient thicknesses (4-16 cm of PMMA). Images from fluoroscopy (low (FL), medium and high) and cine (CI) modes have been archived in DICOM format. Signal-to-noise ratio (SNR), figure of merit (FOM) and high-contrast spatial resolution (HCSR) have been computed from the images. The ratio between the maximum and the minimum value of ESAK per frame for a given fluoroscopy mode between the five systems ranges from 2 to 5 and from 14 to 38 for CI mode. SNR, FOM and HCSR showed a great variability for the different acquisition modes (AMs) and PMMA thickness. In the near future, it is urgent to upgrade Chilean legislation on radiation protection to incorporate quality assurance programmes that will allow us to evaluate and optimise the X-ray systems used in medical applications. Increments in doses per frame when increasing phantom thickness and when used CI runs instead of FL runs can be considered by the cardiologist in the good management of patient dose and allow them to select the best imaging AM during clinical procedures.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Radiological Investigation I: X-ray and CT