Hemodynamic efficacy of sequential hemoclip application using the Olympus HX-110/610 reloadable clipping device in
Anke Ende1, Yurdaguel Zopf, Roland Heide
1Department of Medicine I, University of Erlangen-Nuremberg, Erlangen, Germany.
Background:
Hemoclip application in GI-hemorrhage has proven to be effective. Clinical experience shows that multiple clips are frequently necessary. In 2005, an easily reloadable clip-applicator was introduced. We evaluated the hemodynamic efficacy of this new device.
Material/Methods:
We prospectively compared the new clipping device (Olympus HX 110/610) in a validated experimental setting using the compactEASIE®-simulator for GI bleeding. The artificial blood circulation system in the simulator was connected to a pressure transducer. Four investigators of different endoscopic experience (1000-6000 endoscopies) treated 12 bleeding sources each, with up to 6 clips for each bleeding location. Pressures were recorded to objectify the additive effects of sequential clip application on the reduction in vessel diameter. The intervention was abandoned if a maximum measurable pressure of 300 mmHg was achieved.
Results:
Hemoclip application led to a significant increase of peak pressure (91±100 mmHg, p<0.001) and mean pressure (95±99 mmHg, p<0.001), representing a significant reduction in vessel diameter. Pooled data showed a significant stepwise increase in mean and maximum system pressure, resulting in reduction of vessel diameter up to the fifth hemoclip. On average, 5 clips (range 1-6) were used. More experienced endoscopists achieved a higher increase in mean pressure (167 and 118 mmHg vs 72 and 23 mmHg, p<0.05). Mean reloading time was 39 seconds (19-49 sec).
Conclusions:
Sequential application of multiple hemoclips led to an increasing effect, comparable to the results of previous clinical trials. The number of hemoclips applied correlated inversely, but not significantly, with the endoscopist´s experience. Expensive single-use clips appear dispensable in view of the short reloading time.
Related Concept Videos
Extrinsic and Intrinsic Pathways of Hemostasis
The Extrinsic Pathway
The extrinsic pathway of coagulation is typically initiated by tissue damage that exposes blood to tissue factor (TF), a protein released by the damaged tissue cells outside the blood vessels—this interaction with TF triggers biochemical reactions involving specific clotting factors. The key player here is Factor VII, which forms a...
Introduction to Hemostasis
The three phases of hemostasis involve many clotting factors present in plasma and several substances released by platelets and injured tissue cells. It is a fast, localized, and...
Anticoagulant Drugs: Low-Molecular-Weight Heparins