Closed-loop insulin delivery during pregnancy complicated by type 1 diabetes
Helen R Murphy1, Daniela Elleri, Janet M Allen
1Institute of Metabolic Science, Metabolic Research Laboratories, University of Cambridge, Hills Road, Cambridge, UK. hm386@medschl.cam.ac.uk
Objective:
This study evaluated closed-loop insulin delivery with a model predictive control (MPC) algorithm during early (12-16 weeks) and late gestation (28-32 weeks) in pregnant women with type 1 diabetes.
Research Design And Methods:
Ten women with type 1 diabetes (age 31 years, diabetes duration 19 years, BMI 24.1 kg/m(2), booking A1C 6.9%) were studied over 24 h during early (14.8 weeks) and late pregnancy (28.0 weeks). A nurse adjusted the basal insulin infusion rate from continuous glucose measurements (CGM), fed into the MPC algorithm every 15 min. Mean glucose and time spent in target (63-140 mg/dL), hyperglycemic (>140 to ≥ 180 mg/dL), and hypoglycemic (<63 to ≤ 50 mg/dL) were calculated using plasma and sensor glucose measurements. Linear mixed-effects models were used to compare glucose control during early and late gestation.
Results:
During closed-loop insulin delivery, median (interquartile range) plasma glucose levels were 117 (100.8-154.8) mg/dL in early and 126 (109.8-140.4) mg/dL in late gestation (P = 0.72). The overnight mean (interquartile range) plasma glucose time in target was 84% (50-100%) in early and 100% (94-100%) in late pregnancy (P = 0.09). Overnight mean (interquartile range) time spent hyperglycemic (>140 mg/dL) was 7% (0-40%) in early and 0% (0-6%) in late pregnancy (P = 0.25) and hypoglycemic (<63 mg/dL) was 0% (0-3%) and 0% (0-0%), respectively (P = 0.18). Postprandial glucose control, glucose variability, insulin infusion rates, and CGM sensor accuracy were no different in early or late pregnancy.
Conclusions:
MPC algorithm performance was maintained throughout pregnancy, suggesting that overnight closed-loop insulin delivery could be used safely during pregnancy. More work is needed to achieve optimal postprandial glucose control.
Related Concept Videos
Diabetes Mellitus: Type 2 and Gestational
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Pathophysiology of Diabetes
Type 1 diabetes is characterized by autoimmune-mediated destruction of pancreatic β cells, with environmental factors potentially triggering this process in genetically susceptible individuals. Despite many not having a family history, certain genes increase susceptibility, suggesting a...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into rapid-acting...
Diabetes Mellitus: Introduction