Related Experiment Video
Updated: Jun 5, 2026

Autoradiography as a Simple and Powerful Method for Visualization and Characterization of Pharmacological Targets
Published on: March 12, 2019
Evaluation of a new phosphor plate technology for neonatal portable chest radiographs
Mervyn Cohen1, Donald Corea, Matthew Wanner
1Department of Radiology, Riley Hospital for Children, Indianapolis, IN 46202, USA. mecohen@iupui.edu
Rationale And Objectives:
The aim of this study was to evaluate a new thick-needle phosphor plate for computed radiography.
Materials And Methods:
Two studies were performed. Patients acted as their own controls. In the first study, old powder and new needle phosphor plate technologies were compared. Twenty infants were identified who had undergone chest x-rays with both systems within 3 days of each other. Exposure factors were constant. In the second study, standard and reduced exposure techniques (tube current-time product reduced by 20%) using the needle phosphor technology were compared. Twenty babies who had been imaged with both standard and reduced exposures within 3 days of each other were evaluated.
Results:
There was a significant preference for images obtained with the new needle phosphor technology compared with the older powder technology (P < .01). Using the new needle phosphor plates, a dose reduction of 20% could be achieved without a significant detectable difference between the high-dose and low-dose images (P < .19).
Conclusions:
For the study comparing old and new phosphor plate technology at fixed exposure, images were better with the new technology. Using the new plate technology, dose can be decreased by ≥20%.
More Related Videos
06:46Pulsed Laser Diode-Based Desktop Photoacoustic Tomography for Monitoring Wash-In and Wash-Out of Dye in Rat Cortical Vasculature
Published on: May 30, 2019
06:15Protocol and Guidelines for Point-of-Care Lung Ultrasound in Diagnosing Neonatal Pulmonary Diseases Based on International Expert Consensus
Published on: March 6, 2019
Related Concept Videos
Photoluminescence: Applications
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Determination of Crystal Structures
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...