Related Experiment Video
Updated: Jun 5, 2026

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Surface cooling after cardiac arrest: effectiveness, skin safety, and adverse events in routine clinical practice
Salam Jarrah1, John Dziodzio, Christine Lord
1Division of Pulmonary and Critical Care Medicine, Maine Medical Center, Portland, ME, USA.
Background:
Effectiveness of cooling and adverse events (AEs) involving skin have not been intensively evaluated in cardiac arrest survivors treated with therapeutic hypothermia (TH) when induced and maintained with a servomechanism-regulated surface cooling system.
Methods:
Retrospective review of sixty-nine cardiac arrest survivor-events admitted from April 2006-September 2008 who underwent TH using the Medivance Arctic Sun Temperature Management System. A TH database and medical records were reviewed, and nursing interviews conducted. Primary endpoint was time from initiation to target temperature (TT; 32-34 °C). Secondary endpoints were cooling rate, percentage of hypothermia maintenance phase at TT, effect of body-mass index (BMI) on rate of cooling, and AEs.
Results:
Mean time to the target temperature (TT) was 2.78 h; 80% of patients achieved TT within 4 h; all did within 8 h. Patients were at TT for 96.7% of hypothermia maintenance; 17% of patients had >1 hourly temperature measurement outside TT range. Mean cooling rate during induction phase was 1.1 °C/h, and was not associated with BMI. Minor skin injury occurred in 14 (20%) patients; 4 (6%) were device-related. Skin injuries were associated with shock (P = 0.04), and decubitus ulcers were associated with left ventricular ejection fraction <45% (P = 0.004). AEs included shivering (94%), hypokalemia (81%), hyperglycemia (57%), pneumonia (23%), bleeding (22%), post-cooling fever (17%), and bacteremia (9%).
Conclusions:
The Arctic Sun Temperature Management System was an effective means of performing therapeutic hypothermia after cardiac arrest. Infrequent skin injuries were associated with vasopressor use and low ejection fraction.
More Related Videos
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
05:43Determination of Continuity Index Values in Atrial Fibrillation Ablation with Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Cardiopulmonary Resuscitation IV: Pharmacological Management
Cardiopulmonary Resuscitation III: AED Use
Cardiopulmonary Resuscitation I: Adult
Decreased Body Temperature
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Cardiac Catheterization IV: Nursing Management