A CT acquisition technique to generate images at various dose levels for prospective dose reduction studies
Katie W Hulme1, John Rong, Beth Chasen
1Department of Imaging Physics, The University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Blvd., Box 1352, Houston, TX 77030, USA.
Objective:
The purpose of this article is to determine whether the average of N CT images acquired at a particular dose (D) has image noise equivalent to that of a single image acquired at a dose of N × D.
Materials And Methods:
An electron density phantom, an image quality phantom, and an adult anthropomorphic phantom were scanned multiple times on a 16-MDCT scanner at five effective tube current-rotation time product (mAs) settings (130 kVp; 12, 24, 48, 72, and 144 mAs). Lower-mAs images were averaged to simulate higher-mAs images. Differences in CT number and image noise between simulated and acquired images were quantified using the electron density phantom. Image quality phantom images were scored by three physicists to investigate differences in low- and high-contrast resolution. A forced-choice observer study was performed with three radiologists using anthropomorphic phantom images to evaluate differences in overall image quality.
Results:
The CT number was, on average, reproduced to within 1 HU, and image noise was reproduced to within 4%, which is below the threshold for visibly perceptible differences in noise. Low- and high-contrast resolution were not degraded, and simulated images were visually indistinguishable from acquired images.
Conclusion:
For the dose range studied, it was concluded that the image quality of a CT image produced by averaging multiple low-mAs CT images is identical to that of a high-mAs image acquired at equivalent effective dose, when all other acquisition and reconstruction parameters are held constant. Prospective CT dose-reduction studies may be feasible by acquiring multiple low-dose scans instead of a single high-dose scan. Simulated high-dose images could be interpreted clinically, whereas lower-dose images would be available for an observer study.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging