Increasing efficiency in the ACR Digital Mammography QC Manual's DBT Z-resolution evaluation
Ryan F Fisher1, Nicholas B Bevins2, Katie W Hulme3
1Department of Radiology, Case Western Reserve University/The MetroHealth System, Cleveland, Ohio, USA.
Journal of Applied Clinical Medical Physics
|May 16, 2026
Summary
Reducing regions of interest (ROIs) in digital mammography quality control (QC) can improve efficiency. A 2-speck method for Z-resolution evaluation yields results similar to the standard 6-speck approach with fewer measurements.
Area of Science:
- Medical Imaging Physics
- Quality Control in Healthcare
- Radiologic Technology
Background:
- The ACR Digital Mammography (DM) QC manual specifies Z-resolution evaluation using pseudo full width half max (pFWHM) derived from speck group signals.
- This evaluation involves measuring up to 63 regions of interest (ROIs) across multiple phantom slices, comparing pFWHM to a baseline.
Purpose of the Study:
- To assess if reducing the number of ROIs in the Z-resolution evaluation impacts the reliability of results.
- To determine if a simplified measurement protocol can maintain diagnostic accuracy in digital mammography QC.
Main Methods:
- Retrospective analysis of Z-resolution data from 468 surveys across 147 mammography units.
- Calculation of pFWHM using the standard 6-speck mean (pFWHM6), single speck (pFWHM1), and 2-speck mean (pFWHM2).
- Evaluation of measurement deviation and failure detection rates for each pFWHM method using longitudinal data.
Main Results:
- The 2-speck method (pFWHM2) showed significantly lower average absolute error (3.9%) compared to the single-speck method (pFWHM1) at 6.0% relative to pFWHM6.
- Longitudinal testing revealed pFWHM2 had a mean absolute deviation of 5.6% versus 7.8% for pFWHM1, closely matching pFWHM6 at 5.1%.
- Both pFWHM1 and pFWHM2 detected component-related failures with 0% false negatives, but pFWHM1 had a higher false positive rate (12.9%) than pFWHM2 (2.8%).
Conclusions:
- A single-speck pFWHM measurement leads to high variability and unreliable Z-resolution evaluations in ACR DM QC.
- The 2-speck pFWHM method provides clinically equivalent results to the 6-speck method while reducing measurements by 57%.
- Implementing the 2-speck method enhances testing efficiency without compromising the accuracy of Z-resolution evaluation.