Related Experiment Videos
In vitro detection of enterococcal vancomycin resistance
1Clinical Microbiology Laboratories, University of Chicago Medical Center, Illinois 60637-1463.
Antimicrobial Agents and Chemotherapy
|September 1, 1990
Summary
Detecting vancomycin resistance in enterococci is challenging. Standard methods like disk diffusion and automated systems may fail to identify low-to-moderate resistance levels, potentially leading to reporting errors.
Area of Science:
- Microbiology
- Clinical Diagnostics
- Antimicrobial Resistance
Background:
- Enterococci are a significant cause of nosocomial infections.
- Vancomycin resistance in enterococci (VRE) is a growing public health concern.
- Accurate detection of VRE is crucial for effective treatment and infection control.
Purpose of the Study:
- To evaluate the performance of different methods in detecting varying levels of vancomycin resistance in enterococcal isolates.
- To compare agar dilution, broth microdilution, disk diffusion, and an automated system (AMS Vitek) for VRE detection.
Main Methods:
- Agar dilution was used as the gold standard method.
- Broth microdilution, disk diffusion, and AMS Vitek were tested against six enterococcal isolates with known vancomycin resistance levels.
- Standard and modified (increased inoculum or prolonged incubation) protocols were employed.
Main Results:
- Low-level vancomycin resistance (MIC 16-32 µg/mL) was only detected by agar and broth dilution.
- Moderate-to-high level resistance (MIC 128-256 µg/mL) was missed by disk diffusion and AMS Vitek under standard conditions.
- Modified protocols (10-fold increased inoculum) improved detection for disk diffusion and AMS Vitek for moderate-to-high resistance.
- High-level resistance (MIC 2048 µg/mL) was detected by all methods.
Conclusions:
- The ability of different methods to detect vancomycin resistance in enterococci varies significantly.
- Standard disk diffusion and automated systems may lead to underestimation of VRE, especially for low-to-moderate resistance.
- Variations in detection capabilities can complicate VRE incidence monitoring and lead to critical susceptibility reporting errors.