Serratia marcescens outbreak in a neonatal intensive care unit related to the exit port of an oscillator
Tracy M Macdonald1, Joanne M Langley, Tim Mailman
1Department of Infection Prevention and Control Services, IWK Health Center, Halifax, Canada.
Objective:
To interrupt transmission of Serratia marcescens colonization in a neonatal intensive care unit and determine the source of ongoing transmission.
Design:
Multidisciplinary outbreak investigation and simulation of droplet generation by a high-frequency oscillator using fluorescent dye.
Setting:
Level III neonatal intensive care unit.
Patients:
Very low birth weight premature infants with respiratory failure.
Interventions:
Infection control interventions, pulsed-field gel electrophoresis of isolates to determine relatedness, and construction of a scavenging system to capture the circuitry condensate expelled by the oscillator exit port.
Measurements And Main Results:
Affected infants were housed in the same geographic site. Serratia marcescens isolates were indistinguishable or closely related using pulsed-field gel electrophoresis. Fluorescent droplet splatter from the circuitry, generated when no containment device covered the exit valve, was visible up to 49 in (107.8 cm) from the source.
Conclusions:
Implementation and adherence to infection control measures is essential to prevent transmission of opportunistic pathogens among ventilated infants. Oscillators can generate droplets that travel farther than 1 m from the source.
Insights
A neonatal intensive care unit outbreak of Serratia marcescens was investigated. Infection control measures and containment of oscillator circuitry condensate are crucial to prevent transmission of this opportunistic pathogen.
Area of Science:
- Infectious Disease Epidemiology
- Neonatal Care
- Microbiology
Background:
- Neonatal intensive care units (NICUs) are vulnerable to healthcare-associated infections.
- Serratia marcescens is an opportunistic pathogen that can cause outbreaks in vulnerable populations.
- Ventilated premature infants are at high risk for respiratory infections.
Purpose of the Study:
- To identify the source and interrupt transmission of Serratia marcescens colonization in a NICU.
- To evaluate the role of medical equipment in pathogen dissemination.
Main Methods:
- A multidisciplinary outbreak investigation was conducted in a Level III NICU.
- Pulsed-field gel electrophoresis (PFGE) analyzed Serratia marcescens isolates for relatedness.
- Droplet generation from a high-frequency oscillator was simulated using fluorescent dye.
Main Results:
- Affected infants were cohoused, suggesting localized transmission.
- PFGE revealed indistinguishable or closely related Serratia marcescens isolates.
- Uncontained oscillator circuitry expelled droplets up to 107.8 cm, demonstrating potential for wide dissemination.
Conclusions:
- Strict adherence to infection control protocols is vital for preventing pathogen transmission in NICUs.
- Medical equipment, such as oscillators, can aerosolize and transmit pathogens over significant distances.
- Containment systems for medical device condensate are necessary to mitigate infection risk.
Related Concept Videos
Healthcare Associated Infections I: Iatrogenic, Exogenic and Endogenic
HAIs significantly increase the cost of health care. Extended stays in healthcare institutions, increased disability, increased costs of medications, including specialized antibiotics, and prolonged recovery times add to the patient's expenses and the healthcare institution and funding bodies. Common...
Bacterial Meningitis
Investigation of Disease Outbreaks
Bacterial Meningitis II: Pathophysiology
Botulism
Bacterial Meningitis I: Introduction