Related Experiment Video
Updated: Jun 4, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Surface roughness of femoral head prostheses after dislocation
Kenny T Mai1, Christopher Verioti, Darryl D'Lima
1Hanford Community Medical Center, Hanford, CA, USA. kennymai@hanfordortho.com
Abstract:
The effects of damaged femoral heads on long-term wear in total hip arthroplasties are not well known. In the study reported here, we compared the surface roughness of dislocated femoral heads, retrieved at time of revision, with that of heads revised for reasons other than dislocation. The dislocated heads, including 6 cobalt-chrome (Co-Cr), 2 oxidized zirconium, and 2 alumina (ceramic) heads, were compared with nondislocated Co-Cr and ceramic heads. Scratch marks on the dislocated Co-Cr and alumina heads were considerably smaller and shallow than those on the dislocated oxidized zirconium heads. Mean surface roughness of the dislocated heads was 368 nm (Co-Cr), 376 nm (alumina), and 2137 nm (oxidized zirconium). On the contrary, the mean surface roughness for nondislocated Co-Cr and alumina heads was 307.44 nm (outlier excluded) and 138.8 nm, respectively. Our data suggest that increased surface damage and roughness can occur after dislocation.
