[Determination of dimethyl sulphate in environmental samples with GC-ECD]
Xuan-fu Qin1, Xia-li Zou, Yuan-qian Li
1Department of Sanitary Technology, West China School of Public Health, Sichuan University, Chengdu 610041, China.
Objective:
To develop an gas chromatography-electrophoretic capture detection (GC-ECD)method for determining dimethyl sulphate (DMS) in environmental samples.
Methods:
The environmental samples were dissolved with acetone and the DMS was reacted with KI. The reaction product of methyl iodide was extracted with Hexane and determined by GC-ECD, with HP-5 (30.0 m x 0.32 mm x 0.25 microm) as column, high purity nitrogen as carries, 60 degrees C for column temperature and 240 degrees C for ECD and injector temperature, and 5:1 for split ratio.
Results:
The linear range appeared from 0.05 microg/mL to 2.0 microg/mL. The detection limit was 0.0011 microg/mL. The precision [measured by relative standard deviation (RSD)] of peak area and reserved time for the standard solution was 4.95% and 0.15% respectively for intra-day detections, and 5.99% and 0.83% respectively for inter-day detections. Satisfactory results were obtained for the analysis of environmental samples. The RSD for the water and soil samples were 9.6% and 7.5% respectively and the recoveries were 76.0%-85.0% and 82.0%-110.0% respectively.
Conclusion:
The method was simple, accurate, sensitive and applicable for DMS analysis in environmental samples.
More Related Videos
11:06GC-based Detection of Aldononitrile Acetate Derivatized Glucosamine and Muramic Acid for Microbial Residue Determination in Soil
Published on: May 19, 2012
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...