Related Experiment Video
Updated: Jun 4, 2026

Molecular Beam Mass Spectrometry With Tunable Vacuum Ultraviolet (VUV) Synchrotron Radiation
Published on: October 30, 2012
Characterization of bioparticles using a miniature cylindrical ion trap mass spectrometer operated at rough vacuum
Zhiqiang Zhu1, Caiqiao Xiong, Gaoping Xu
1Key Laboratory of Analytical Chemistry for Living Biosystems, Institute of Chemistry Chinese Academy of Sciences, Beijing, 100190, China.
Abstract:
A miniature cylindrical ion trap mass spectrometer (CIT-MS) equipped with an inexpensive mechanical pump was constructed, and used to measure the masses of cells and microparticles generated by laser induced acoustic desorption ionization. Compared with a previous lab scale quadrupole ion trap mass spectrometer (QIT-MS), the novel miniature CIT-MS had smaller volume (the radius r(0)=5 mm), simpler ion trap fabrication and overall instrument construction, required a lower trapping voltage, and reduced the weight, power and cost of the instrument. The CIT-MS was calibrated using standard polystyrene beads of 2.982 µm diameter. The CIT-MS was used to measure the total dry weight of human red blood cells (RBCs) from a healthy female adult (2.12×10(13) Da) and a patient with anemia (1.35×10(13) Da). The coefficient of variance (CV) for the healthy individual was 21% and that for the patient was 30.4%. The CIT-MS was also applied to guinea pig RBCs and the total dry weight was determined as 1.34×10(13) Da with a CV of 37.9%. These measurements are consistent with previous QIT-MS measurements. The new miniaturized instrument has potential for applications in field-portable, biological and aerosol analysis.
Related Concept Videos
Mass Analyzers: Common Types
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometers
Mass Analyzers: Overview
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
High-Resolution Mass Spectrometry (HRMS)
