Introduction and clinical evaluation of servo-controlled applanation resonance tonometry
Gauti Jóhannesson1, Per Hallberg, Anders Eklund
1Department of Clinical Sciences, Ophthalmology, Umeå University, Umeå, Sweden. gauti.johannesson@ophthal.umu.se
Purpose:
In recent years, Applanation Resonance Tonometry (ART) has been suggested for intraocular pressure (IOP) measurements. The manual version of ART (ART(manual) ) has been further developed, and to improve usability, an automatic servo-controlled prototype (ART(servo) ) has been proposed. The aim of this study was to assess the limits of agreement (LoA) of ART(manual) and ART(servo) as compared with the reference method, Goldmann Applanation Tonometry (GAT).
Methods:
This was a prospective single-centre study on 152 eyes from 77 subjects. It was designed according to International Standard Organization's (ISO) requirements for tonometers (ISO 2001). Intraocular pressure was measured six times/method in a standardized order. The ART technique has two available analysis procedures: a dynamic one that measures IOP during the indentation phase and a static one that causes a Goldmann-like measurement during two seconds of full applanation. The 95% LoA was defined as ±1.96 × standard deviation of difference against GAT.
Results:
Mean IOP for GAT was 19.1 mmHg (range: 10-37 mmHg). The 95% LoA of ART(manual) was ±4.5 mmHg for both dynamic and static analyses. The 95% LoA of ART(servo) was ±5.7 mmHg for dynamic and ±4.9 mmHg for static analyses.
Conclusions:
This study confirms that the ART methodology is feasible. The further developed ART(manual) fulfilled the ISO standard with both the dynamic and the static analysis techniques. ART(servo) with static analysis was close to fulfilling the standard but failed to do so in the highest IOP range. ART(servo) has the potential to greatly improve usability if further development is completed.
More Related Videos
06:15Anterior High-Resolution Optical Coherence Tomography in the Diagnosis and Therapeutic Monitoring of Ocular Surface Squamous Neoplasia
Published on: August 9, 2024
09:47A Methodological Protocol and Considerations for Transcranial Ultrasonic Stimulation in Exploratory Clinical Human Studies
Published on: December 12, 2025
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessment of blood pressure in brachial artery(two-step method)