Measurement of the 109Cd half-life
R Van Ammel1, S Pommé, J Paepen
1European Commission, Joint Research Centre, Institute for Reference Materials and Measurements, Retieseweg 111, Geel, Belgium. raf.van-ammel@ec.europa.eu
Abstract:
The half-life of (109)Cd was measured by following the decay of sources from a radiochemically pure solution with two different measuring systems: an ionisation chamber and a high-purity germanium (HPGe) detector. The measurements were performed over a period of 3.6 years, i.e. about 2.8 half-lives of (109)Cd. The resulting half-life values and detailed uncertainty budgets (k=1) are presented for both systems. The result obtained with the ionisation chamber, 462.36 (33) days, and the one obtained with the HPGe detector, 461.92 (76) days, are mutually consistent. The weighted mean of our measured values, T(1/2)((109)Cd)=462.29 (30) days, is consistent with the currently recommended values of 461.4 (12) days (Schönfeld and Dersch, 2004; IAEA, 2007) and 462.0 (3) days (Xiaolong et al., 2010). From a set of selected experimental values published after 1970, a "partially weighted mean" (Pommé and Spasova, 2008) of T(1/2)((109)Cd)=462.36 (39) days was calculated. More measurements are needed to resolve the discrepancies among literature data and to reduce the final uncertainty on the (109)Cd half-life.
Related Concept Videos
Radioactive Decay and Radiometric Dating
Half-life of a Reaction
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution
Measurement of Bioavailability: Pharmacodynamic Methods
![Automated Preparation of [68Ga]Ga-3BP-3940 on a Synthesis Module for PET Imaging of the Tumor Microenvironment](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F68356.jpg&w=3840&q=50)