Related Experiment Videos
The strength of a posterior element claw at one versus two spinal levels
J W Roach1, R B Ashman, R N Allard
1Research Department, Texas Scottish Rite Hospital, Dallas 75219.
Abstract:
In the Cotrel-Dubousset and Texas Scottish Rite Hospital spinal implants systems, the upper hook on the convex rod is usually clawed to prevent posterior pull out. A transverse process hook and an upwardly directed pedicle hook are usually inserted on the same lamina level. However, clawing two levels instead of one is a simpler surgical procedure. We biomechanically compared the posterior pull-out strength of a one-level versus two-level construct. The average load of failure of the double-level construct was significantly greater (p less than 0.05) as compared with the single level. Thus, the two-level construct is not only easier to insert, it is also stronger.