Accuracy of the Microlife large-extra large-sized cuff (32-52 cm) coupled to an automatic oscillometric device
Serena Masiero1, Francesca Saladini, Elisabetta Benetti
1Department of Clinical and Experimental Medicine, University of Padova, Padua, Italy.
Abstract:
To determine the accuracy of the large-extra large-sized (L-XL) cuff (32-52 cm) coupled to a Microlife WatchBP Office ABI blood pressure measuring device tested according to the requirements of the International Protocol of the European Society of Hypertension. The L-XL cuff tested in this study is designed to provide accurate blood pressure measurements in patients with large arms (arm circumference≥32 cm) over a wide range of arm circumferences using a single 145±1×320±1 mm bladder. The evaluation was made in 33 patients with a mean±standard deviation age of 53±17 years (range: 30-96 years). Their systolic blood pressure (SBP) was 142±21 mmHg (range: 110-180 mmHg), diastolic blood pressure (DBP) was 87±14 mmHg (range: 62-106 mmHg) and arm circumference was 36±5 cm (range: 32-50 cm). Blood pressure measurements were made in the sitting position. The L-XL cuff coupled to the WatchBP Office ABI passed all three phases of the European Society of Hypertension protocol for SBP and DBP. Mean blood pressure differences between device and observer were -1.3±5.1 mmHg for SBP and -1.8±5.8 mmHg for DBP. Similar device-observer differences were observed in patients divided into two subgroups according to whether their arm circumference was above or below the median in the group. These results indicate that the L-XL cuff coupled to the WatchBP Office ABI monitor provides accurate blood pressure readings in patients with large arms over a wide range of arm circumferences.
More Related Videos
06:04Pulse-Wave Velocity, Flow-Mediated Dilation, and Carotid Intima-Media Thickness to Assess Cardiovascular Risk in Population with Metabolic Syndrome
Published on: September 27, 2024
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: