Related Experiment Video
Updated: Jun 4, 2026

04:41
Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
Proximal humeral replacement using a fixed-fulcrum endoprosthesis
D Griffiths1, P D Gikas, C Jowett
1Royal National Orthopaedic Hospital, Brockley Hill, Stanmore, Middlesex HA7 4LP, UK. dylgrif@hotmail.com
The Journal of Bone and Joint Surgery. British Volume
|March 2, 2011
Summary
Proximal humerus tumor replacement using a fixed-fulcrum endoprosthesis shows reliable outcomes. A new constrained liner significantly reduced dislocation rates, improving patient results.
Area of Science:
- Orthopedic Surgery
- Oncology
- Biomedical Engineering
Background:
- Tumors of the proximal humerus require surgical intervention, often necessitating complex reconstructive procedures.
- Massive endoprosthetic replacement is a common surgical option for proximal humerus tumor resection.
- Previous endoprosthetic designs faced challenges with instability and dislocation.
Purpose of the Study:
- To evaluate the long-term outcomes of proximal humerus tumor replacement using a fixed-fulcrum massive endoprosthesis.
- To assess the efficacy of a novel constrained humeral liner in reducing dislocation rates.
- To compare functional outcomes between the original prosthesis and the modified prosthesis with the constrained liner.
Main Methods:
- A retrospective review of 68 consecutive patients who underwent proximal humerus tumor replacement between 1997 and 2007.
- Assessment of functional outcomes using the Musculoskeletal Tumor Society (MSTS) Score and the Toronto Extremity Salvage Score (TESS).
- Analysis of complication rates, specifically focusing on prosthetic dislocation, and comparison between the original and constrained liner designs.
Main Results:
- The study included 42 surviving patients with a mean follow-up of 5 years and 11 months.
- The mean MSTS score was 72.3% and the mean TESS score was 77.2% for the overall cohort.
- The original prosthesis had a dislocation rate of 25.9%; however, no dislocations occurred in the four patients who received the new constrained humeral liner.
Conclusions:
- Endoprosthetic replacement for proximal humerus tumors using the fixed-fulcrum prosthesis is a reliable procedure with good functional outcomes.
- The introduction of a constrained humeral liner significantly mitigates the risk of dislocation, enhancing prosthesis performance.
- This improved prosthesis design offers a reliable solution for proximal humerus tumor reconstruction, minimizing major complications.