Measuring the relative concentration of H2 and D2 in HD gas with gas chromatography
C Steven Whisnant1, Patrick A Hansen, Travis D Kelley
1James Madison University, Harrisonburg,Virginia 22807, USA. whisnacs@jmu.edu
Abstract:
Frozen-spin polarized hydrogen deuteride (HD) targets are being prepared for photonuclear experiments at Jefferson Lab. The targets are frozen HD gas. To polarize and achieve the required spin-lattice relaxation times, the targets are made from high purity HD gas in which the residual H(2) and D(2) concentrations are known. To determine these residual concentrations, a gas chromatograph is used. The separation of the gas components is done in a column cooled to ≈120 K by immersing it in a mixture of liquid nitrogen and isopentane and using neon as the carrier gas. The different hydrogen isotopes have different transit times through the column and their arrival is registered by a thermal conductivity detector. The peaks in the chromatograms are fit using an exponentially modified Gaussian line shape to extract the areas of these peaks. The ratios of areas are corrected for differences in thermal conductivity to give the relative concentrations. Here, H(2) and D(2) concentrations on the order of a few percent have been measured with uncertainties of less than ± 5%. The linear response of the system is confirmed to within about ± 6.3%.
More Related Videos
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
Published on: December 10, 2015
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
¹H NMR of Labile Protons: Deuterium (²H) Substitution