Introduction And Objectives:
Sprint Fidelis defibrillation leads are prone to early failure. Most of the reported series come from a single institution. This paper describes the clinical experience in nine Spanish hospitals.
Methods:
Clinical, implant, and follow-up data of all patients with a Sprint Fidelis lead were analyzed. All cases of lead failure were identified, medium-term lead survival was calculated, and possible predictors for lead failure were determined.
Results:
In total, 378 leads in 376 patients were studied. The mean age (male 85.7%) was 64.9 ± 13.6 years. The majority of patients (59.8%) had ischemic heart disease. Mean left ventricular ejection fraction was 33.4% ± 14.5%. Left subclavian vein puncture was used in 74.8%. During a mean follow-up of 30.9 ± 14 months, 16 lead failures have occurred, with a lead survival of 96.1% at 36 months after implantation. Eleven of 16 lead failures were caused by failure of pace/sense conductors, 3 by defects in the high-voltage conductor, and 2 by defects in both types of conductors. A less depressed left ventricular ejection fraction was associated with an increased probability of lead failure (42.4% ± 16% vs. 33% ± 14.3%; P =.011). Three hospitals presented a rate of lead failure higher than 10%; the rate was less than 5% in the remaining 6 hospitals.
Conclusions:
In this multicenter series of 378 leads, the 3-year estimated survival was higher than that reported in prior series. Clinical presentation of lead failures was similar to that reported previously. Left ventricular ejection fraction and hospital of implantation were variables associated to lead failure.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Dysrhythmias V: Evaluating Dysrhythmias
Cardiac Catheterization III: Left Heart Catheterization
Dysrhythmias I: Introduction
Cardiac Catheterization II: Right Heart Catheterization
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...