Correction methodology for the spectral interfering γ-rays overlapping to the analytical peaks used in the analysis
1Institute of Nuclear Sciences, Ankara University, Besevler Yerleskesi, Tandoğan, Turkey. haluk.yucel@ankara.edu.tr
Abstract:
In the γ-ray spectrometric analysis of the radionuclides, a correction factor is generally required for the spectral interfering γ-rays in determining the net areas of the analytical peaks because some interfering γ-rays often might contribute to the analytical peaks of interest. In present study, a correction methodology for the spectral interfering γ-rays (CSI) is described. In particular, in the analysis of (232)Th contained in samples, the interfering γ-rays due to (226)Ra, (235)U, (238)U and their decay products often overlap to the peaks of interest from (232)Th decay products, and vise versa. For the validation of the proposed CSI method, several certified reference materials (CRM) containing U and Th were measured by using a 76.5% efficient n-type Ge detector. The required correction factors were quantified for spectral interference, self-absorption and true coincidence summing (TCS) effects for the relevant γ-rays. The measured results indicate that if one ignores the contributions of the interfering γ-rays to the analytical peaks at 583.2 keV of (208)Tl and 727.3 keV of (212)Bi, this leads to a significantly systematic influence on the resulted activities of (232)Th. The correction factors required for spectral interference and TCS effects are estimated to be ∼13.6% and ∼15.4% for 583.2 keV peak. For the 727.3 keV peak, the correction factor is estimated to be ∼15% for spectral interference, and ∼5% for the TCS effects at the presently used detection geometry. On the other hand, the measured results also indicate that ignoring the contribution of the interfering γ-rays to the areas of the analytical peaks at 860.6 keV of (208)Tl, 338.3 and 911.2 keV of (228)Ac does not lead to any significant systematic influence on the (232)Th analysis. Because these factors are remained generally less than ∼5%, i.e., within overall uncertainty limits. The present study also showed that in view of both the spectral interference and TCS effects, the "cleaner" analytical peaks are seen at 338.3 keV (11.25%) and 911.2 keV (26.13%) of (228)Ac when high resolution γ-rays spectrometry was used in the (232)Th activity measurements. Therefore, they can be adopted as the "reference" peaks in the (232)Th analysis.
Related Concept Videos
Atomic Emission Spectroscopy: Interference
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
NMR Spectrometers: Resolution and Error Correction
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...