Methodologies for the analytical determination of ferrate(VI): a review
Zhiyong Luo1, Meagan Strouse, Jia-Qian Jiang
1College of Chemistry and Chemical Engineering, Chongqing University, Chongqing, China.
Abstract:
This paper reviewed different analytical techniques to determine concentrations of the environmentally-friendly oxidant ferrate(VI) (Fe(VI)O(2-)(4), Fe(VI)) anion in solid and solution samples. The techniques included volumetric (chromite and arsenite), electrochemical (cyclic voltammetry and potentiometry), and spectrophotometric methods (Fourier transform infrared, Mossbauer, UV-Visible, 2,2-azino-bis(3-ethylbenzothiazoline-6-sulfonate (ABTS), and fluorescence). The importance of the techniques were briefly described. This paper also presented a newly developed method for the determination of low concentrations (0.25 μM) of Fe(VI)) using the reaction between Fe(VI) and iodide (I(-)), which formed the yellowish color of I(-)(3) that could be measured spectrophotometrically at 351 nm. The increase in I(-)(3) was linear with respect to Fe(VI) added (0.25-76 μM) in buffered solutions at pH 5.5-9.3. The linearity was not affected by the ions present in tap water.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides and...
Voltammetry: Factors Affecting Measurements
Complexometric Titration: Overview