Related Experiment Videos
Validation of the Welch Allyn SureBP (inflation) and StepBP (deflation) algorithms by AAMI standard testing and BHS
1Department of Pediatrics, University of Tennessee Health Science Center, Memphis, 38105, USA. bsalpert@uthsc.edu
Objectives:
We evaluated two new Welch Allyn automated blood pressure (BP) algorithms. The first, SureBP, estimates BP during cuff inflation; the second, StepBP, does so during deflation. We followed the American National Standards Institute/Association for the Advancement of Medical Instrumentation SP10:2006 standard for testing and data analysis. The data were also analyzed using the British Hypertension Society analysis strategy.
Methods:
We tested children, adolescents, and adults. The requirements of the American National Standards Institute/Association for the Advancement of Medical Instrumentation SP10:2006 standard were fulfilled with respect to BP levels, arm sizes, and ages. Association for the Advancement of Medical Instrumentation SP10 Method 1 data analysis was used.
Results:
The mean±standard deviation for the device readings compared with auscultation by paired, trained, blinded observers in the SureBP mode were -2.14±7.44 mmHg for systolic BP (SBP) and -0.55±5.98 mmHg for diastolic BP (DBP). In the StepBP mode, the differences were -3.61±6.30 mmHg for SBP and -2.03±5.30 mmHg for DBP. Both algorithms achieved an A grade for both SBP and DBP by British Hypertension Society analysis.
Conclusion:
The SureBP inflation-based algorithm will be available in many new-generation Welch Allyn monitors. Its use will reduce the time it takes to estimate BP in critical patient care circumstances. The device will not need to inflate to excessive suprasystolic BPs to obtain the SBP values. Deflation is rapid once SBP has been determined, thus reducing the total time of cuff inflation and reducing patient discomfort. If the SureBP fails to obtain a BP value, the StepBP algorithm is activated to estimate BP by traditional deflation methodology.
Related Concept Videos
Testing a Claim about Standard Deviation
The hypothesis testing for the claim of population standard deviation (or variance) requires the data and samples to be random and unbiased. The population distribution also must be normal. There is no specific requirement on the sample size as the estimation is based on the chi-square distribution.
As a first step, the hypothesis (null and alternative) concerning the claim about...
Assessment of blood pressure in brachial artery(two-step method)
Behrens–Fisher Test
This test is...