Calibration of KVp meter used in quality control tests of diagnostic X-ray units
Ngoc Toan Tran1, Takeishi Iimoto, Toshiso Kosako
1Department of Nuclear Engineering and Management, School of Engineering, The University of Tokyo, 7-3-1 Hongo, Bunkyo, Tokyo 113-8656, Japan. tntoanvn@gmail.com
Abstract:
In diagnostic radiology, the tube voltage [peak kilovoltage (kV(p))] is one of the most important parameter affecting both radiation exposure and image contrast. So, an accurate kV(p) meter is necessary to control kV(p) in the medical radiography practice with the overall uncertainty less than ± 5 % according to IEC 61676. Therefore, both invasive and non-invasive calibration methods of kV(p) meter were established and applied to different kinds of commercial quality control instruments for diagnostic radiology. Calibration of kV(p) meter by the invasive method is the most accurate (with uncertainty of 1.67 %, k=2); however, the non-invasive method also provides good results (with uncertainty of 3.12 %, k=2). Due to their detailed design, the commercial kV(p) meters have various responses with X-ray beam, so the working regime of a particular device type must be appropriately selected with a specific X-ray machine used for calibration of kV(p) meter.
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality Assurance
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Constant Volume Calorimetry