Thermal marks as a signal processing aid for a portable capillary electropherograph
Andrus Seiman1, Merike Vaher, Mihkel Kaljurand
1Department of Chemistry, Tallinn University of Technology, Akadeemia tee 15, Tallinn, Estonia. andrusseiman@gmail.com
Abstract:
The interpretation of raw signals in capillary CE can be challenging if there are unknown peaks, or the signal is corrupt due to baseline fluctuations, EOF velocity drift, etc. Signal processing could be required before results can be interpreted. A suite of signal processing algorithms has been developed for CE data analysis, specifically for use in field experiments for the detection of nerve agents using portable CE instruments. Everything from baseline correction and electropherogram alignment to peak matching and identification is included in these programs. Baseline correction is achieved by interpolating a new baseline according to points found using all local extremes, by applying an appropriate outliers test. Irreproducible migration times are corrected by compensating for EOF drift, measured with the aid of thermal marks. Thermal marks are small disturbances in the capillary created by punctual heating that move with the velocity of EOF. Peaks in the sample electropherogram are identified using a fuzzy matching algorithm, by comparing peaks from the sample electropherogram to peaks from a reference electropherogram.
Related Concept Videos
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...