Providing PEEP during neonatal resuscitation: which device is best?
Jennifer A Dawson1, Angela Gerber, C Omar F Kamlin
1Neonatal Services, The Royal Women's Hospital, Melbourne, Australia.
Aim:
The study aims to compare three commonly used neonatal resuscitation devices, the Laerdal self-inflating bag with a positive end expiratory pressure (PEEP) valve, a T-piece resuscitator (T-piece) and a flow-inflating bag to provide peak inflation pressure (PIP) and PEEP.
Methods:
Participants were asked to use each device to give positive pressure ventilation to a modified neonatal mannequin via a face mask to achieve 40-60 inflations per minute, aiming for a PIP/PEEP of 30/5 cm H₂O. A manometer was visible to participants with each device. PIP, PEEP, percentage leak at the face mask and expired tidal volume were measured using a hot-wire anemometer. We analysed 20 inflations from each participant for each device.
Results:
Fifty participants provided PIP and PEEP with each device. The T-piece was the most accurate and consistent. The flow-inflating bag had the most variation. The leak was lowest with the self-inflating bag and PEEP and highest with the flow-inflating bag, but all had wide variation.
Conclusion:
Each device was able to provide PIP and PEEP when used appropriately. When compared with other resuscitation devices, the T-piece provided the most accurate and consistent PIP and PEEP.