Co-60 tomotherapy: a treatment planning investigation
1Department of Medical Physics, Saskatoon Cancer Centre, Saskatoon, Saskatchewan S7N 4H4, Canada. pat.cadman@saskcancer.ca
Purpose:
The authors investigate the plan quality and treatment times that may be achieved with Co-60 tomotherapy delivery for clinical IMRT cases.
Methods:
A research version of PINNACLE treatment planning system software (V9.1) enabled the authors to specify custom source profiles for modeling of cylindrical Co-60 sources. The calculated profiles were validated against measurements for simulated MLC leaf openings. The reduction in dose due to a partially obscured source was analyzed. The thread effect was investigated for a source of typical linac spot dimensions and 2.0 and 2.8 cm diameter cylindrical Co-60 sources. Co-60 tomotherapy plans for three clinical treatment sites--prostate, brain, and head-and-neck--were generated for the Co-60 sources and compared to linac-based segmental IMRT plans in terms of the DVHs produced. Treatment times were also determined.
Results:
The custom source profile utility allowed the authors to obtain good agreement between calculated and measured profiles for simulated MLC leaf openings with the commercial Co-60 source. It was found that the thread effect is significantly reduced for Co-60 sources and is not a clinical concern even for the large slice width (4.8 cm) and pitch value (0.5) studied. Co-60 tomotherapy plans for three clinical treatment sites compared favorably to the original segmental IMRT plans in terms of the DVHs produced. Treatment times, comparable to the actual segmental IMRT treatment times, may be achieved for a high activity Co-60 source and dual-slice delivery may reduce these times further.
Conclusions:
It may be possible to achieve clinically viable treatment times with Co-60 tomo-therapy delivery without unacceptable loss of plan quality in terms of the DVHs produced.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...