Design, implementation and validation of a motorized wedge filter for a telecobalt machine (Bhabhatron-II)
Rajesh Kumar1, D C Kar, S D Sharma
1Radiological Physics & Advisory Division, Bhabha Atomic Research Centre, CTCRS Building, Anushaktinagar, Mumbai-400094, India. raj_resh@rediffmail.com
Abstract:
A universal wedge filter of 15W × 20 cm(2) and 60° nominal wedge angle is designed and placed between the collimating jaws and penumbra trimmers inside the treatment head. A pneumatically driven actuating mechanism toggles the wedge between the wedge IN position and wedge OUT position. The effective wedge angles were determined using an analytical formula. An accumulated wedge profile at a depth of 10 cm which was measured using a 2D profiler and dose values at depths of 10 cm and 20 cm for the same experimental setup were used as input parameters in the formula used for determining effective wedge angles. The relationship between the wedge beam weight and effective wedge angle was established. The planned wedge angles were compared with the measured wedge angles and the differences are found to be less than 2° throughout the range of field sizes. Planned doses for various field sizes and wedge angles were measured for verification and the differences were found to be less than 1.8%. This study established that the relationship between the beam weights and effective wedge angles implemented for the motorized wedge filter of medical linacs is not directly applicable for the motorized wedge filter of Telecobalt.
Related Concept Videos
Design Example
Wedges
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...
Design of Transmission Shafts
Electro-mechanical Systems
A key component of the DC motor is the armature, a rotating circuit positioned within a magnetic field. As an electric current passes through the...