Related Experiment Video
Updated: Jun 2, 2026

Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
Establishing an alternative method for the quantitative analysis of polychlorinated dibenzo-p-dioxins and
Jayne de Vos1, Peter Gorst-Allman, Egmont Rohwer
1National Metrology Institute of South Africa, Lynnwood Ridge 0040, South Africa. bjdevos@nmisa.org
Abstract:
Comprehensive Gas Chromatography-Time-of-Flight Mass Spectrometry (GC×GC-TOFMS) methodology has been refined for the analysis of polychlorinated dibenzo-p-dioxins (PCDDs) and polychlorinated dibenzofurans (PCDFs) in samples with different matrices. This is specifically for application in developing countries where access to gas chromatography-high resolution mass spectrometry (GC-HRMS) and highly skilled personnel is limited. The method, using an Rxi-5 Sil MS column in the first dimension ((1)D) coupled with an Rtx-200 column in the second dimension ((2)D), was used to quantify PCDDs and PCDFs in different environmental sample matrices. The results were compared with those obtained using GC-HRMS and good agreement was observed. The limit of detection (LOD) for the method (300fg on column for spiked soil samples) was determined using an Rxi-XLB ((1)D) column coupled with an Rtx-200 column ((2)D). Preliminary South African sample results are also discussed. Isomer specificity for different tetrachloro dibenzo-p-dioxins (TCDDs) and tetrachloro dibenzofurans (TCDFs) was investigated using a commercial standard. Adequate resolution was achieved. The method as described has great attraction for developing countries being both financially and operationally favourable.
More Related Videos
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
11:44Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
Published on: March 6, 2016
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Detectors-II