Validation of the Welch Allyn ProBP 3400: a device for modern medical practice
1Department of Pediatrics, University of Tennessee Health Science Center, Memphis, Tennessee 38103, USA. bsalpert@uthsc.edu
Objectives:
As medical care evolves, automated sphygmomanometers must be able to provide accurate and timely blood pressure (BP) data to the healthcare provider. We evaluated the performance of a new hand-held, battery-operated device, the Welch Allyn ProBP 3400, to the AAMI SP 10 criteria. We also performed British Hypertension Society data analysis.
Methods:
A total of 111 (97 adults, 14 children) participants were tested using blinded and paired observers performing auscultation of Korotkoff (K1 and K5) sounds.
Results:
Ninety-two participants had acceptable data, giving 275 paired data sets for systolic BP (SBP) and diastolic BP (DBP) analyses. The mean ± standard deviation of the observer/device differences were -1.1 ± 7.5 mmHg for SBP and -2.1 ± 6.0 mmHg for DBP. These passed the AAMI SP10 requirements. The ProBP 3400 achieved an A grade for both SBP and DBP.
Conclusion:
The ProBP 3400 passed the AAMI standard requirements and achieved A grade for both SBP and DBP. The device is less in cost and completely portable. It transmits BP data directly into the electronic medical record using a Universal Serial Bus connection. These features will improve the efficiency and accuracy of BP data available to clinicians, thus improving the quality of care.
Related Concept Videos
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Automated Microbial Diagnostics