Measuring the Complete-arch Distortion of an Optical Dental Impression
Published on: May 30, 2019
Evaluation of the accuracy of different transfer impression techniques for multiple implants
Júlio César Brigolini de Faria1, Laís Regiane Silva-Concílio, Ana Christina Claro Neves
1Department of Prosthodontics, University of Taubaté, Taubaté, SP, Brazil.
Brazilian Oral Research
|May 4, 2011
Summary
The study compared implant impression techniques. Splinting square copings (SCS) demonstrated superior accuracy and fidelity, closely matching the control group, making it a clinically acceptable method for transferring implant positions.
Area of Science:
- Dental Implantology
- Prosthodontics
- Materials Science in Dentistry
Background:
- Accurate impression techniques are crucial for successful dental implant restorations.
- Existing methods for transferring implant positions can introduce inaccuracies.
- Evaluating novel techniques is essential for improving clinical outcomes.
Purpose of the Study:
- To assess and compare the accuracy of three distinct implant transfer impression techniques.
- To determine the most precise method for capturing implant positions for superstructure fabrication.
Main Methods:
- Four groups were established: tapered copings without splint (TC), square copings without splint (SC), square copings splinted with floss and resin (SCS), and a control group (CG) master model.
- Polyether material and individual trays were used for impressions.
- Image analysis software (UTHSCSA ImageTool) quantified gap discrepancies between the master superstructure and cast models; statistical analysis included ANOVA and Student's t-test.
Main Results:
- The control group (CG) exhibited the lowest mean gap (2.03 ± 0.00 µm).
- The splinted square copings (SCS) group showed the smallest gap among tested techniques (6.51 ± 0.09 µm), with no statistical difference from the CG.
- Tapered copings (TC) and non-splinted square copings (SC) presented significantly larger gaps (14.74 ± 3.41 µm and 12.08 ± 2.56 µm, respectively) compared to the CG.
Conclusions:
- The splinted square copings (SCS) technique demonstrated high accuracy and fidelity in transferring implant positions, comparable to the master model.
- All evaluated techniques yielded clinically acceptable gap values.
- The SCS technique is recommended for its precision in implant impression procedures.