Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method
Heiko Gesche1, Detlef Grosskurth, Gert Küchler
1Institute of Vegetative Physiology, Charité-Universitätsmedizin Berlin, Berlin, Germany.
Pulse transit time (PTT) can estimate systolic blood pressure (SBP) using pulse wave velocity (PWV). While PTT shows correlation with SBP, individual calibration is needed for accurate, absolute blood pressure measurements.
Area of Science:
- Biomedical Engineering
- Cardiovascular Physiology
Background:
- Pulse transit time (PTT) and pulse wave velocity (PWV) are established indirect measures correlated with systolic blood pressure (SBP).
- Non-invasive blood pressure monitoring remains a significant challenge in clinical practice.
Purpose of the Study:
- To develop and validate a non-linear function for determining absolute SBP from PWV using a single-point calibration.
- To assess the reliability of PTT-based SBP measurement against traditional cuff-based methods.
Main Methods:
- A non-linear algorithm was developed using PWV data from 13 volunteers to establish a SBP-PWV relationship.
- The derived function was applied to 50 volunteers, with individual offset correction, to calculate SBP.
- PTT was measured via ECG R-spike to finger plethysmography; reference SBP was obtained using a sphygmomanometric aneroid device.
Main Results:
- A significant correlation (r = 0.83) was observed between cuff-measured SBP and PTT-derived SBP across 276 measurements in 50 volunteers.
- Individual correlation coefficients ranged from r = 0.69 to r = 0.99.
- Bland-Altman analysis revealed limits of agreement of ±19.8 mmHg, indicating notable differences between methods.
Conclusions:
- PTT-based SBP measurement shows significant correlation with cuff-based SBP.
- Relevant differences exist between PTT and cuff methods, partly due to measurement variability during exercise.
- Absolute SBP can be measured using PTT with individual offset correction for the BP-PWV relationship.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure in the Leg
Preparation: