In vitro and in vivo comparison of wrist MR imaging at 3.0 and 7.0 tesla using a gradient echo sequence and identical
Jurek A Nordmeyer-Massner1, Michael Wyss, Gustav Andreisek
1Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
Purpose:
To evaluate in vivo MR imaging of the wrist at 3.0 Tesla (T) and 7.0T quantitatively and qualitatively.
Materials And Methods:
To enable unbiased signal-to-noise ratio (SNR) comparisons, geometrically identical eight-channel receiver arrays were used at both field strengths. First, in vitro images of a phantom bottle were acquired at 3.0T and 7.0T to obtain an estimate of the maximum SNR gain that can be expected. MR images of the dominant wrist of 10 healthy volunteers were acquired at both field strengths. All measurements were done using the same sequence parameters. Quantitative SNR maps were calculated on a pixel-by-pixel basis and analyzed in several regions-of-interest. Furthermore, the images were qualitatively evaluated by two independent radiologists.
Results:
The quantitative analysis showed SNR increases of up to 100% at 7.0T compared with 3.0T, with considerable variation between different anatomical structures. The qualitative analysis revealed no significant difference in the visualization of anatomical structures comparing 3.0T and 7.0T MR images (P>0.05).
Conclusion:
The presented results establish the SNR benefits of the transition from 3.0T to 7.0T for wrist imaging without bias by different array designs and based on exact, algebraic SNR quantification. The observed SNR increase nearly reaches expected values but varies greatly between different tissues. It does not necessarily improve the visibility of anatomic structures but adds valuable latitude for sequence optimization.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...