Correction of frequency drifts induced by gradient heating in 1H spectra using interleaved reference spectroscopy
Thomas Lange1, Maxim Zaitsev, Martin Buechert
1Department of Radiology, Medical Physics, University Hospital Freiburg, Freiburg, Germany. thomas.lange@uniklinik-freiburg.de
Purpose:
To evaluate the impact of heating-induced frequency drifts on single-voxel spectroscopy and to demonstrate correction strategies based on the interleaved reference scan technique.
Materials And Methods:
Frequency drifts induced by gradient heating are assessed for two clinical 3 Tesla (T) whole body MR systems. The interleaved reference scan (IRS) method is used for correcting these frequency drifts in 1H spectra in vitro and in vivo. For severely drift-affected spectroscopy experiments, a feedback-based version of the IRS sequence is proposed, which adds the functionality of a frequency lock to prevent a degradation of the water suppression.
Results:
It is shown that the line widths of the spectral resonances can be largely reduced with the interleaved reference scan method, resulting in considerably improved peak resolution and signal-to-noise ratio. The feedback-based IRS method additionally allows for stable water suppression, even in the presence of very strong frequency drifts.
Conclusion:
If spectroscopy scans are combined with imaging scans with a high gradient duty cycle such as diffusion-weighted imaging or functional MRI, a drift correction with IRS can considerably improve the validity of data analysis in research studies.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are slanted or...
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature from...
Atomic Emission Spectroscopy: Interference
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...