Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Tonometry in keratoconic eyes before and after riboflavin/UVA corneal collagen crosslinking using three different
Maria G Gkika1, Georgios Labiris, Vassilios P Kozobolis
1Eye Institute of Thrace and Department of Ophthalmology, Democritus University of Thrace, Alexandroupolis, Greece. mggkika@yahoo.gr
Purpose:
To evaluate the Goldmann applanation tonometer (GAT), the Pascal dynamic contour tonometer (PDCT), and the ocular response analyzer (ORA) tonometer in measuring intraocular pressure (IOP) in keratoconic eyes before and after riboflavin/ultraviolet A corneal collagen crosslinking (CXL), to assess agreement among devices and to analyze the impact of some ocular parameters on their measurements.
Methods:
Fifty keratoconic eyes were included. Intraocular pressure was measured with GAT, PDCT, and ORA before and after CXL. Fifty nonkeratoconic eyes served as controls. Device agreements were calculated by Bland-Altman analysis. The effect of some ocular characteristics on IOP measurement differences between tonometers was determined.
Results:
Between the 2 groups, there were statistically significant differences in all examined parameters. Preoperatively, in both groups a statistically significant difference was found in IOP measurements among devices (p<0.05). Bland-Altman analysis showed a bias among devices. On average, PDCT overread GAT and ORA. The IOP measurement differences were better predicted by corneal resistance factor. Postoperatively, in keratoconus eyes, there was no statistically significant difference in IOP measurements among the 3 tonometers (p>0.05). The IOP readings with all tonometers after treatment were higher than those obtained preoperatively; however, not to a statistically significant level, with the exception of PDCT and ORA readings at the first month postoperatively. Corneal resistance factor measurements have no significant change after CXL.
Conclusions:
Pascal dynamic contour tonometer could provide more consistent and closer to the true IOP readings than GAT and ORA in healthy eyes with corneal thickness outside the 520-550 µm range, in keratoconus patients and after CXL. Corneal resistance factor was associated significantly with agreement among devices.