Analytical comparison of three quantitative immunochemical fecal occult blood tests for colorectal cancer screening
Lydia Guittet1, Elodie Guillaume, Romuald Levillain
1INSERM ERI3 Cancers & Populations, University of Caen Basse-Normandie (UCBN), University Hospital (CHU de Caen), Caen, France. guittet-l@chu-caen.fr
Background:
The superiority of several immunochemical fecal occult blood tests (I-FOBT) over guaiac-based tests in colorectal cancer screening is now established. The aim of this study was to compare the analytical performance of 3 quantitative I-FOBTs.
Methods:
Stool samples from 10 healthy volunteers, initially I-FOBT negative, supplemented with human blood, were used to compare reproducibility and stability of measurement at varying storage temperatures (4°C, 10°C, 20°C, and 30°C) and durations before test analysis (1 to 10 days) for 3 I-FOBTs (New Hemtube/Magstream HT, OC-Auto sampling bottle3/OC-Sensor DIANA, and FOB Gold/SENTiFOB). Concentrations ranging from 0 to 350 μg Hb/g of feces were evaluated.
Results:
The measurement reproducibility of OC-Sensor was superior to Magstream and far superior to FOB Gold. For all tests, variability was essentially related to sampling. Detected hemoglobin (Hb) levels were substantially lower for all tests at temperatures above 20°C. At 20°C, this loss in concentration was less important with OC-Sensor (significant 1.7% daily decrease vs. 7.4% for Magstream and 7.8% for FOB Gold). At 30°C, daily loss was 8.6% with OC-Sensor, whereas after 24 hours, only 30% of the original Hb was detected with FOB Gold, compared to 70% with Magstream. No Hb was detected on day 5 for the latter 2 tests.
Conclusions:
About reproducibility and temperature stability, OC-Sensor performed better than Magstream and far better that FOB Gold.
Impact:
Independently of the chosen test, the delay between sampling and test processing should be reduced, the maximal admissible delay depending on ambient temperature.
More Related Videos
04:57Comparative Analysis of Automatic Fecal Analyzer versus Direct Wet Smear Microscopy for Detecting Parasitic Infections in Stool Samples
Published on: April 25, 2025
15:17Colorectal Cancer Cell Surface Protein Profiling Using an Antibody Microarray and Fluorescence Multiplexing
Published on: September 25, 2011
Related Concept Videos
Serum Laboratory Studies, Stool Test, Breath Test
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...