Evaluation of a pulse oximeter sensor tester
Shuba Dugani1, Iljaz Hodzovic, Seema Sindhakar
1Department of Anaesthetics, Intensive Care and Pain Medicine, Cardiff University, UK.
Objective:
The Lightman is intended to test the optical and electrical properties of a pulse oximeter probe including the wavelength of the light emitting diode by means of a micro spectrometer. The aim of this study was to evaluate the ability of the Lightman to detect faulty pulse oximeter finger probes by testing the accuracy of the wavelength of the light emitting diode in isolation from the monitor.
Methods:
The pulse oximeter measurements of arterial oxygen saturation from the "accurate" and "inaccurate" probes, as identified by the Lightman, were compared with arterial saturation determined by a co-oximeter. Data was analysed from 63 sets of measurements. In addition, we conducted a national survey to determine the testing procedures used by the Biomedical Engineering departments to evaluate the accuracy of pulse oximeter devices.
Results:
The bias [95% limits of agreement] for accurate, over-reading and under-reading probes were 0.17% [3.6 to -3.3], 1.44% [5.4 to -2.5] and -1.6% [2.6 to -5.8] respectively. The response rate to the national survey was 75% (142/189); a pulse oximeter tester was used by 93/142 (65%) trusts.
Conclusions:
Our findings suggest that the Lightman can detect faulty probes and predict reasonably accurately the direction of the probe's error. The Lightman may be considered as a useful tool to assess the accuracy of pulse oximeters. The national survey highlighted a wide variation in the testing procedure utilised to evaluate the accuracy of pulse oximeters. Introduction of guidelines regarding the testing procedure would promote a uniform practice.
More Related Videos
10:45A Community-based Stress Management Program: Using Wearable Devices to Assess Whole Body Physiological Responses in Non-laboratory Settings
Published on: January 22, 2018
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Special considerations while measuring pulse
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...