A bench study of the effects of leak on ventilator performance during noninvasive ventilation
Yoshitoyo Ueno1, Nobuto Nakanishi, Jun Oto
1Department of Emergency and Critical Care Medicine, University of Tokushima Graduate School, Tokushima, Japan.
Background:
During noninvasive ventilation (NIV), leak around the mask may cause inadequate ventilatory support or patient-ventilator asynchrony such as auto-triggering. Some NIV ventilators may be better than others at compensating for leak.
Methods:
We bench studied 3 NIV ventilators (Vision, Carina, and Trilogy100) and 2 ICU ventilators (Puritan Bennett 840 and Evita XL) to assess how they coped with 2 leak levels and zero leak during NIV. With a 2-bellows-in-a-box lung model we simulated spontaneous breathing with tidal volumes of 300 mL and 500 mL, at pressure support of 0 and 10 cm H(2)O and PEEP of 5 and 10 cm H(2)O. We affixed the airway opening of the lung model to the mouth of a mannequin head and secured a mask on the mannequin face. We created a medium leak and a large leak with different size holes, and measured PEEP in the presence of leak. We also measured the actual pressure-support values and calculated the deviations from the set pressure-support value and the pressure-time product (PTP) of the airway opening pressure below and above baseline.
Results:
With the medium leak only the Vision and Carina maintained the set PEEP and pressure support. With the large leak the pressure support was decreased with all the tested ventilators. With the larger leak and pressure support of 10 cm H(2)O the PTP below baseline for triggering increased with 2 ventilators, and the PTP above baseline for supporting the patient's inspiratory effort decreased with all 5 ventilators. The larger tidal volume increased the PTP below baseline with all 5 ventilators and at all leak sizes.
Conclusions:
Some of the ventilators compensated for leak better than others. With the larger leak none of the ventilators maintained the set PEEP or pressure support.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...