Mechanical stability of endotracheal tube fixation: a performance comparison between two tube holders and a conventional method
Related Experiment Video
Updated: Oct 31, 2026

The Rigid Tube as an Alternative in Controlling the Problematic Airway
Published on: June 6, 2020
Alireza Abootalebi Ghahnavieh1, Mehran Haddadi1, Mehdi Nasr Isfahani2,3
1Department of Emergency Medicine, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran.
Background:
Accidental extubation remains a significant clinical concern in airway management, particularly in emergency and anesthetic settings. While commercial tube holders have shown improved fixation stability over adhesive tape, cost and availability limit their use in many healthcare systems. This study introduces a newly developed Iranian endotracheal tube (ETT) holder, Persifix, and evaluates its mechanical performance relative to a commercial Locktite model and the conventional adhesive tape method.
Methods:
A total of 170 controlled tests were conducted using a human airway manikin. Three fixation methods-adhesive tape, Persifix, and Locktite tube holders-were assessed under lateral and frontal force vectors using weights of 750 g and 1000 g. Displacement was measured in both depth and lateral dimensions under dynamic (tug) and static (non-tug) conditions. Statistical analysis was performed using Kruskal-Wallis and Mann-Whitney U tests, with Bonferroni's post hoc comparisons (α = 0.05).
Results:
The adhesive tape method demonstrated significantly greater displacement in depth across most test conditions (e.g., left side, 1000 g, no tug: 2.54 ± 3.63 mm vs. Persifix: 1.76 ± 1.78 mm, p = 0.032; Locktite: 0.53 ± 0.44 mm, p < 0.001). In frontal force tests with 1000 g and tug, the Persifix showed reduced displacement (0.41 ± 0.16 mm) compared to both adhesive tape (2.62 ± 4.25 mm, p = 0.001) and Locktite (0.52 ± 1.33 mm, p = 0.032). Lateral displacement in width was also significantly lower with both tube holders (e.g., 750 g with tug: Persifix 3.34 ± 5.20 mm vs. adhesive tape 3.16 ± 15.00 mm, p = 0.005). No statistically significant differences were observed between the Persifix and Locktite tube holders in most conditions (p > 0.05), indicating comparable fixation stability.
Conclusion:
The Iranian tube holder, Persifix, demonstrated statistically significant reductions in ETT displacement compared to adhesive tape across multiple force vectors and weight conditions. Its mechanical performance was comparable to the commercial Locktite model, suggesting its potential utility as a cost-effective and clinically reliable alternative for airway stabilization in resource-limited settings.
More Related Videos
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes:
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...