Related Experiment Video
Updated: Jun 1, 2026

Leveraging Turbidity and Thromboelastography for Complementary Clot Characterization
Published on: June 4, 2020
[Establishment of normal reference values for thromboelastography on Chinese population in Beijing]
Hong-wen Ji1, Li Ma, Xu-rong Gao
1Department of Transfusion, Fuwai Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College, National Center for Cardiovascular Diseases, Beijing 100037, China. jihongwen@hotmail.com
Objective:
To determine the normal values for thromboelastography (TEG) in Chinese healthy adult volunteers residing in Beijing for over three years and compare them with those of the manufacturer's.
Methods:
A total of 137 healthy adult volunteers were enrolled from June 2010 to August 2010. The technique was standardized with citrated blood and kaolin activator. And a Haemoscope 5000 device was employed. The TEG parameters analyzed were R, K, α, maximal amplitude (MA), LY30 and coagulation index (CI). All volunteers underwent the tests of prothrombin time (PT), activated partial thromboplastin time (APTT), thrombin time (TT) and plasma fibrinogen level with the same blood sample.
Results:
The reference ranges of 95% for 137 volunteers were R: 3.8 - 8.4 min, K: 0.8 - 3.3 min, α-Angle: 46.2 - 76.2°, MA: 50.0 - 70.8 mm, LY30: -3.3% - 4.0% and CI: -3.8 - 2.9. Overall, 24.1% (33/137) of the volunteers had at least one abnormal parameter while 7.3% (10/137) would have been considered coagulopathy had the manufacturer's reference values been used, resulting in a test specificity of 76.0%. As compared with the western ethnicity (the manufacturer's reference values), Chinese healthy volunteers were associated with lower fibrinogen functions. There were significantly different in R, K, α-Angle, MA and CI between men and women groups (all P < 0.01).
Conclusion:
This study supports the manufacturer's recommendation that each institute should determine its own normal reference values.