Characterization of airway device cuff volumes at simulated altitude
Jennifer Law1, Aaron Bair, Jason Capra
1University of California Davis Medical Center, Sacramento, CA, USA. jelaw@utmb.edu
Introduction:
Cuff volume of an air-filled airway device varies inversely with ambient pressure at altitude. This may result in problems with ventilation, aspiration, and tissue ischemia in intubated patients transported by aircraft. We aimed to characterize cuff volume changes in airway devices as a function of altitude.
Methods:
Four inflatable airway devices lendotracheal tube (ETT), Combitube, King tube, and laryngeal mask airway (LMA)I, each inflated with air or water, were evaluated for cuff volume changes in an altitude chamber simulating ascent and descent from ground level to 15,000 ft (4572 m). A novel cuff-less supraglottic device called i-gel was also tested. Multivariate linear regression was used to assess the effect of altitude and cuff content on cuff volume.
Results:
We found a linear relationship between air-filled cuff volume and altitude in all the inflatable airway devices. The Combitube (correlation coefficient R = 0.94) and King tube (R = 0.98) showed the clearest linear relationship; the ETT (R = 0.70) and LMA (0.86) showed modest correlations. With water-filled cuffs, the rate of increase was relatively smaller in all the inflatable devices except the ETT, which remained constant. The difference between air- and water-filled cuffs was statistically significant in each inflatable device (P < 0.001). The i-gel showed no volume change at any of the tested altitudes.
Conclusions:
Cuff volume of inflatable airway devices increased linearly with altitude. The dual-cuffed supraglottic devices showed greater volume changes, likely due to the combined effect of their two cuffs. Of the inflatable devices, the water-filled ETT was the least sensitive to altitude changes.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
08:44Dual Test Gas Pulmonary Diffusing Capacity Measurement During Exercise in Humans Using the Single-Breath Method
Published on: February 2, 2024
Related Concept Videos
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Respiratory Volumes and Capacities I
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Respiratory Volumes and Capacities
Cardiopulmonary Resuscitation II: ACLS Airway Management
Lung Capacity