Related Experiment Video
Updated: Jun 1, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
[Pacemaker/ICD lead extraction with the excimer laser sheath: a single center experience]
1Veterans Administration Medical Center, University of Minnesota, Mineapolis, MN, USA. jianmingli924@yahoo.com
Excimer laser-assisted lead extraction is a safe and effective method for removing pacemaker and implantable cardioverter-defibrillator leads. This study confirms its utility in complex cases with high success rates.
Area of Science:
- Cardiology
- Medical Devices
- Interventional Procedures
Background:
- Pacemaker and implantable cardioverter-defibrillator (ICD) leads can require extraction due to infection, malfunction, or upgrades.
- Lead extraction can be challenging, especially for older or complex leads, posing risks of complications.
Purpose of the Study:
- To evaluate the safety and effectiveness of the Excimer Laser Sheath Lead Removal System for extracting cardiac leads.
- To assess the outcomes of lead extraction in patients with various underlying cardiac conditions and lead characteristics.
Main Methods:
- A retrospective analysis of 16 patients who underwent lead extraction using the Excimer Laser Sheath at a single center.
- Data collected included patient demographics, cardiac function, lead characteristics, indications for extraction, and complications.
Main Results:
- Thirty-one leads were successfully extracted in 16 patients, with the Excimer Laser Sheath used in 25 leads.
- The procedure demonstrated a high success rate with no major complications reported.
- Patients often had severe heart disease, with a mean ejection fraction of 36.4%.
Conclusions:
- The Excimer Laser Sheath Lead Removal System is a safe and effective tool for pacemaker and ICD lead extraction.
- This technology facilitates lead removal even in patients with complex cardiac histories and lead types.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization